DETERMINATION OF EFFECT OF A LEAD FILTER ON THE AVERAGE ENERGY OF PHOTON RADIATION OF A MEDICAL LINEAR ELECTRON ACCELERATOR
Abstract and keywords
Abstract (English):
The article considers the change in the average energy of photon radiation with the upper limit of the spectrum of 6 MeV, generated by a medical linear electron accelerator (or LINAC for short), when radiation passes through a lead filter of various thicknesses. The calculation of the spectrum and average energy of photon radiation is carried out using Monte Carlo modelling. Additionally, the coefficients of attenuation of photon radiation by a lead filter are calculated. Particular attention is paid to the influence of the filter on the shape of the spectrum. The validation of the calculation is determined by comparing the model and experimentally obtained attenuation coefficients for kerma.

Keywords:
Monte Carlo modelling, medical linear electron accelerator, kerma power attenuation coefficient, FLUKA
Text
Publication text (PDF): Read Download
References

1. Martynyuk Yu.N., Nurlybaev K., Revkov A.A. Dozimetriya impul'snogo izlucheniya//ANRI. 2018. №№1 (92). S. 1-11.

2. IEC TS 63050:2019: Radiation protection instrumentation - Dosemeters for pulsed fields of ionizing radiation.

3. ISO 4037-1:2019: Radiological protection - X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy. Part 3. Calibration of area and personal dosemeters and the measurement of their response as a function of energy and angle of incidence.

4. J. Karzmak, «A Primer on Theory and Operation of Linear Accelerators in Radiation Therapy», Medical Physics Pub Corp., 56 p., 2018.

5. Dozimetr DKS-AT5350/1 URL: https://atomtex.com/ru/oborudovanie-dlya-kalibrovki-i-poverkidozimetry-etalonnye/dozimetr-dks-at53501 (data obrascheniya: 20 dekabrya 2022 g.).

6. Fluka-4 Manual. URL: https://flukafiles.web.cern.ch/manual/index.html (data obrascheniya: 20 dekabrya 2022 g.).

7. B.J. Giuseppe, B.K. Till, C.A. Francesco, W.C. Pik, E.Q. Salvatore, F.O. Alberto, F.M. Alfredo, «Overview of the FLUKA code», Annals of Nuclear Energy, vol. 82, no. 1. pp. 10-18, 2015.

8. Varian High Energy Linear Accelerators Comparison Chart URL: https://www.oncologysystems.com/resources/linear-accelerator-guides/varian-high-energy-linear-accelerators-comparison-chart (data obrascheniya: 20 dekabrya 2022 g.).

9. B. Bednarz, G. Xu, «Monte Carlo modeling of a 6 and 18 MV Varian Clinac medical accelerator for in-field and out-of-field dose calculations: development and validation», Phys. Med. Biol., vol. 54, no.№4., pp. 43-57, 2009.

10. R. Mohan, K. Jayesh, R. Joshi, «Dosimetric evaluation of 120-leaf multileaf collimator in a Varian linear accelerator with 6-MV and 18-MV photon beams», J. Med. Phys., vol. 33, no. 3. pp. 114-118, 2008.

11. Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients from 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest. URL:https://www.nist.gov/pml/x-ray-mass-attenuation-coefficients (data obrascheniya: 20 dekabrya 2022 g.).

12. H.J. Choi, H. Park, C.Y. Yi, «Determining the energy spectrum of clinical linear accelerator using an optimized photon beam transmission protocol», J. Med. Phys., vol. 46, no. 3, pp. 3285-3297, 2019.

Login or Create
* Forgot password?