CDTE DETECTOR INSTRUMENTAL SPECTRUM CORRECTION IN THE LOW-ENERGY X-RAY REGION
Abstract and keywords
Abstract (English):
The article devotes to algorithm for the instrumental spectrum correction of a CdTe detector in the low-energy X-ray region with a cutoff energy < 150 keV. By minimizing the influence of the detector’s response function, the algorithm allows the instrumental spectrum to be reconstructed, approaching its shape closer to the true one. Particular attention is paid to the low-energy region with the boundary energy up to 100 keV. For a given region, the algorithm makes it possible to correct the spectrum taking into account the difference between the theoretical efficiency curve and the modeled efficiency curve. The main goal of developing an improved algorithm is to determine the characteristics of the radiation quality of an X-ray facility.

Keywords:
Monte Carlo simulation, spectrum correction, CdTe detector, FLUKA
Text
Publication text (PDF): Read Download
References

1. J. Franc, E. belas, R. Grill, P. Hlidek, P. Moravec, «CdTe and CdZnTe crystals for room temperature gamma-ray detectors», Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 434, no. 1, pp. 146-151, 1999.

2. X-123CdTe Complete X-Ray & Gamma Ray Spectrometer. URL: https://www.amptek.com/products/x-ray-detectors/cdte-x-ray-and-gamma-ray-detectors/cdte-x-ray-and-gamma-ray-detector. (data obrascheniya: 20.12.2021).

3. S. Miyajima, «Thin CdTe detector in diagnostic x-ray spectroscopy», Medical Physics, vol. 30, no. 5, pp. 771-777, 2003.

4. S. Miyajima, K. Imagawa, M. Matsumoto, «CdZnTe detector in diagnostic x-ray spectroscopy», Medical Physics, vol. 29, no. 7, pp. 1421-1429, 2002.

5. URL: https://flukafiles.web.cern.ch/manual/fluka.html (data obrascheniya: 02.01.2022.

6. Zagorodnyuk A.A., Korneev S.V. Modelirovanie spektrov standartnyh obrazcov urana U3O8 SRM-969 izmerennyh pri pomoschi poluprovodnikovogo germanievogo detektora//Izvestiya Nacional'noy akademii nauk Belarusi. Seriya fiziko-tehnicheskih nauk. 2017. N 3. C. 92-100.

7. A. Tomal, J. Santos, P. Costa, A.H. Lopez, M.E. Poletti, «Monte Carlo simulation of the response functions of CdTe detectors to be applied in x-ray spectroscopy», Applied Radiation and Isotopes, vol. 100, no. 1, pp. 32-37, 2015.

8. J.L. Campbel, M. Leonard, H. Theodore, P. Tibor, «Simulations of Si(Li) x-ray detector response», X-ray Spectrometry, 2001. vol. 30, no. 1, pp. 230-241.

9. K. Zahraman, M. Roumie, A. Raulo, N. Aruicchio, M. Ayob, «Study of the Thickness of the Dead Layer Below Electrodes, Deposited by Electroless Technique, in CdTe Nuclear Detectors», IEEE Transactions on Nuclear Science, vol.53, no. 1, pp. 378-382, 2006.

10. ISO 4037-1:2019: Radiological protection - X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy. Part 1: Radiation characteristics and production methods.

11. U. Ankerhold. «Catalogue of X-ray spectra and their characteristic data. ISO and DIN radiation qualities, therapy and diagnoxtic qualities, unfiltered X-ray spectra», Physikalisch-Technische Bundesanstalt, PTB-Bcricht Dos 34, 2000.